299x Filetype PDF File size 0.44 MB Source: www.phase-technology.com
WAT-70Xi
Crude Oil Analyzer
for wax appearance temperature/cloud point and wax disappearance temperature
from Phase Technology
Speed & Precision Benefits Wax Appearance Test (WAT) and Wax Disappearance Test (WDT)
for Upstream & Midstream A Breakthrough in Crude Oil Testing
DETECTION OF BOTH WAT AND WDT With a long, notable history as the world Phase Technology’s new WAT-70Xi analyzer
Two critical tests for understanding crude leader in developing test methods for cold revolutionizes the upstream and midstream
oil behavior; also determines rate of wax flow properties detection, Phase Technology oil industries with the world’s first and only
deposition & melting. has now extended its capability to include totally automatic process to measure WAT
FASTER THAN ANY OTHER TEST METHOD a critical measurement for crude oil: and WDT of crudes. Based upon ASTM D5773,
Wax Appearance Temperature (WAT). our proprietary optical light scattering
Results in just 20-40 minutes, compared to Also known as cloud point, wax appearance technique detects phase changes with
average test times of several hours for other temperature is the temperature at which a extreme sensitivity and accuracy.
methods. crude oil sample first
TESTS OPAQUE SAMPLES precipitates solid wax Timely Test Results;
Enhanced optical configuration “sees” through as it is being cooled No Set Up or Cleaning
under prescribed test This important scientific
dark samples. conditions. breakthrough means
SELF CLEANING Similarly, Wax that even the darkest,
Automatic solvent flush after each test run. Disappearance most opaque samples
NO MANUAL SET UP Temperature (WDT) can now be easily
is the temperature at tested, with 1.0°C
Simply inject sample directly into analyzer and which the last wax precision. ust load the
begin test run. solids are melted into sample, and the analyzer
SUPERIOR PRECISION liquid during a warming does the rest, with
cycle. tests completed in just
Repeatability of 1.0° C. 20-40 minutes. There’s
GREATER SENSITIVITY An End to Subjective, no time-consuming
Controlled, automatic test method ensures no Tedious Testing manual set-up needed,
subjectivity in reported results. Until now, trying to and cleaning is done
determine crude oil automatically after
INFORMATIVE, REAL-TIME TEST RESULTS WAT or cloud point was each test.
Complete phase diagram (loop) clearly an imprecise, tedious,
illustrates WAT, relative amount of wax and subjective process. Trusted 70Xi
formation and WDT. Various manual methods Analyzer Design
have been tried, but all The new WAT analyzer is
INTUITIVE, EASY-TO-USE INTERFACE are difficult and time- built on the 70Xi series
Full-color, touch-sensitive, 15” high resolution consuming, yielding platform, with time-
screen and one-touch, preset “favorites” results with unacceptably saving, productivity
wide error margins. features included.
11168 Hammersmith Gate Richmond, B.C. Canada V7A 5H8 604.241.9568
www.phase-technology.com
Technical Specifications
WAT-70Xi-Crude Oil Analyzer
TEST METHOD ASTM D5773 (IP 446) Applications
ASTM D2500 (IP 219 / ISO 3015) equivalent
STATED PRECISION Repeatability Wax appearance temperature (WAT) and wax
1.0 ºC disappearance temperature (WDT) can help predict
SAMPLE TEMPERATURE RANGE -30 ºC to 75 ºC the occurrence of wax deposition in crude oils, and
SYSTEM CLEANING Automatic flush cycle; external solvent tank are therefore useful for upstream and midstream
TEST DURATION 20 to 40 minutes petroleum companies.
REQUIRED OPERATOR TIME 15 seconds In oilfield applications, WAT or WDT can assist
SAMPLE SIZE 3.0 mL required in determining optimum levels of wax crystal
DETECTION METHOD Patented Diffusive Light Scattering (DLS) technology modifiers and/or wax deposition inhibitors. WAT is
COOLING SYSTEM Integrated Peltier device cooling system also an indicator of potential crude incompatibility
DISPLAY Full-color, touch-sensitive, 15” high resolution LCD touch screen and a monitor of evolving crude quality. Crude oils
(3) USB A ports for optional peripherals: flash drive, from the same region may have quite different
label printer, barcode scanner, keyboard, mouse; (1) USB B port characteristics, with disparate rates of wax
OUTPUTS (3) RS-232 serial ports for optional peripherals & networking:
external computer, Phase Technology LTB diagnostic software; deposition and dissolution. Changes of location,
(1) dedicated Service port; (1) 10/100Base-T Ethernet (RJ45) port for extraction depth, evolution of time, or even
networking: LIMS, local area network (LAN)
TEMPERATURE MEASUREMENT ºC or ºF (User selectable) methods of production and blending can all be
ALERTS Buzzer for alarms warnings and prompts (User selectable) verified by WAT.
INTERNAL MEMORY Storage up to 5000 test runs For transport of crude oil via pipeline, railway or
AMBIENT OPERATING 10 to 30 °C (50 to 86 °F) tanker, as well as for oil storage, wax crystals may
ROOM TEMPERATURE Extremes not recommended
restrict flow or create a total blockage. WAT and
Unit Length x Width x Height WDT can help define acceptable operability limits
DIMENSIONS 21.5 x 13.25 x 17.5 inches
(W x D x H) 54.6 x 33.7 x 44.5 cm and prevent the downtime and expense associated
Boxed 29 x 23 x 19 inches with cleaning. WAT is a useful tool for the accurate
74 x 58 x 48 cm
WEIGHT Unit 53 lbs / 24 kg prediction of wax deposition in pipelines and
Boxed 62 lbs / 28 kg storage vessels, resulting in great potential savings.
Electrical 90 – 260 VAC, 47 – 63 Hz The design and development of subseas and land
UTILITY REQUIREMENTS 350 watts pipeline systems and implementation of wax
External NONE remediation options benefit from analysis of WAT
Cooler Bath
data.
ANALYZER
SCREENSHOT
Phase Analyzer Company, LTD www.phase-technology.com email: info@phase-technology.com phone: 604.241.9568
no reviews yet
Please Login to review.